
Calibration ECA 450
60 Instruction 0024-9400
3.12 Calibrate CO-HI Sensor
This procedure spans the optional Carbon Monoxide High sensor (4,000 –
80,000 ppm CO) to a known gas level.
Material Required:
• Calibration Gas Fixture (Section 3.2)
• Gas Cylinder: 4,000 ppm CO in air (refer to Section 6.2)
Procedure:
1. Enter Calibration List screen; then choose CO-HI per Section 3.3 to
display the CALIBRATE CO-HI screen. At the conclusion of this step the
pump should be running.
2. Using the Gas Fixture shown in Figure 3-1, attach a CO calibration gas
cylinder to the analyzer’s GAS inlet port.
3. Adjust regulator of calibration fixture for a flow meter indication of
approximately 2 SCFH.
4. Wait until Measured reading on screen stabilizes (approximately
3 minutes). Then use the
∧∨ and < > keys to enter an Applied value that
equals the concentration stamped on the CO gas cylinder.
The calibration range is from 250 to 11,000 ppm. An applied value that is
outside this range will cause the message “Bad Calibration Entry” to
appear.
Typical Calibrate CO-HI Screen During Calibration:
«««CALIBRATE CO-HI««
Measured 3890 ppm
Applied 04000 ppm
5. Press ENTER to calibrate the analyzer’s Measured value to that of the
Applied value; after which the message “Good Calibration” should appear
along with the Calibration List screen being redisplayed.
If the message “Sensor Output Low” or “Sensor Not Calibrated” appears,
refer Section “5.4 Troubleshooting Guide.”
6. Turn off regulator; then remove gas cylinder from calibration fixture.
Komentarze do niniejszej Instrukcji