Bacharach Flue-Gas Sample Conditioning System Dokumentacja Strona 28

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 27
Operation ECA 450
24 Instruction 0024-9400
8. With the analyzer set up per Section 2.4 Performing a Combustion Test,
press ENTER key to begin testing and logging.
«LOG«««(0025)««NGAS«
This display shows that the analyzer is logging and currently saving data
to internal memory location 25. Observe that the memory location
number will increment as data is being saved. Note that if all memory
locations are used before the end of the logging duration, then “(xxxx)”
appears in place of the memory location number, indicating a memory
over-range condition.
«LOG«««(-PC-)««NGAS«
This display shows that the analyzer is logging data, and that the data is
being sent to only a personal computer.
Tip: Logging can be stopped without loosing data that has already been
collected by pressing the ESC key.
9. At the conclusion of the logging process, the message “Logging Complete”
will appear. And if data was being stored in the analyzer’s internal
memory, then the message “Data in memory nnnn to nnnn” will also
appear where nnnn denotes the internal memory locations that
contain the logged data. Press ENTER to leave logging and display the
Combustion Test HOLD screen.
2.12.2 Turning Data Logging Off
Data logging is automatically turned off . . .
at the end of the logging process when the ENTER key pressed.
if the logging process is interrupted by pressing the ESC key.
Note: If the logging process is interrupted, it cannot be resumed without first
turning logging back on per Section 2.12.1.
Przeglądanie stron 27
1 2 ... 23 24 25 26 27 28 29 30 31 32 33 ... 85 86

Komentarze do niniejszej Instrukcji

Brak uwag