ECA 450 Operation
Instruction 0024-9400
2.15 Differential Pressure Measurement
To measure the differential pressure between two confined areas, do the
following:
1. Press PRESSURE key to display the Pressure screen.
««««««PRESSURE««««««
n.nn xxxx
Where: n.nn = current pressure reading
xxxx = pressure units (refer to Section 2.17.2)
2. If desired, the pressure sensor can be zeroed as described below. To skip
the zeroing procedure, proceed to Step 3.
a) With Pressure screen displayed, again press PRESSURE key to
display the Pressure Zero screen.
Pressure Zero
Disconnect Hose
Then Press ENTER
b) Disconnect all hoses from the ∆P REF and PRESSURE connectors.
This step ensures that both of these connectors are sampling the same
pressure.
c) Press ENTER to zero pressure sensor to atmospheric pressure; after
which, reconnect probe hose that was removed in Step 3b.
Reconnect Hose
3. Connect pressure hose assembly P/N 0024-1103 to analyzer’s ∆P REF
connector (see Figure 1-1). Then insert open end of this hose into area
being used as the reference pressure.
4. Insert probe into area whose pressure is to be measured.
5. The Pressure screen is now displaying the differential pressure between
the two areas. Positive readings indicate that the pressure of the area
being measured is higher than the reference area, while negative
readings indicate the pressure is lower than the reference.