
Fyrite
INSIGHT
Plus Manual Calibration and Maintenance
0024-9487 Rev 2 89
Step T-STACK Calibration Procedure
2 If not already done, turn ON the analyzer and display the CALIBRATION
Menu. Note that this requires password validation (see page 55).
3 Use the UP () and DOWN () arrow buttons to highlight T-Stack, and then
press ENTER to display the CALIBRATE TS-ZERO screen.
"Measured" is the current temperature reading. "Applied" is a known
temperature that will be applied for calibration purposes.
4 Set thermocouple simulator to 32° F (0° C), and then use the UP (), DOWN
(), LEFT (), and RIGHT () arrow buttons to enter an Applied value that
exactly equals the setting of the simulator.
Alternatively: Submerge probe tip into an ice-
thermometer, wait several minutes, and then use the UP () and DOWN ()
arrow buttons to enter an Applied value that exactly equals the thermometer
reading.
NOTE: The calibration range is from 32 to 41° F (0 to 5° C). An
attempt to calibrate outside this range will cause the message
“Applied Value High" (or Low) to appear at the bottom of the
screen.
5 Wait until the Measured reading stabilizes, and then press ENTER to calibrate
the TS-Zero Measured value to that of the Applied value, after which the
message “Good Calibration” should briefly appear followed by the CALIBRATE
TS-SPAN screen.
Komentarze do niniejszej Instrukcji