Fyrite
INSIGHT
Plus Manual Operation
0024-9487 Rev 2 77
4.7.4. Stack Temperature Hot Spot Screen
Hot Spot
Component
Description
Hot Spot Graph
• Press RUN/HOLD to start/stop
the hot spot function.
• T-STACK parameter name
shown in graph.
• Limits are determined
automatically.
• Dynamic “current value” pointer indicates real-time value.
• Color of “current value” pointer is based on limit status:
Black (Top): Hottest reading since last “Clear”
Red (Bottom): Current reading
• Ideally, position probe so current reading (bottom) and
highest reading (top) match.
Left Arrow
• Press the LEFT () arrow to go to the Bar Graph Screen.
Right Arrow
• Press the RIGHT () arrow to go to the main RUN/HOLD
Screen.
Clear (F1)
• Press the F1 button to clear the display and restart if in RUN
mode.
Menu (F2)
• Press the F2 button to return to the Main menu.
4.8. Taking Ambient CO Measurements (Siegert Only)
This procedure takes approximately 15 minutes to complete and provides a
minute-by-minute snapshot of CO readings, as well as a “Max CO” value that
represents the highest CO reading measured during the entire 15-minute test.
Results can be saved to memory, downloaded, and/or printed. Use the
following procedure to perform an ambient CO measurement.
Step Example Procedure for Taking Ambient CO Measurements
1
Turn on the instrument in a fresh air environment and wait for
initialization to complete.
2
Verify successful initialization (no errors).