Bacharach Fyrite INSIGHT PLUS Instrukcja Obsługi Strona 93

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 92
Fyrite
INSIGHT
Plus Manual Calibration and Maintenance
0024-9487 Rev 2 87
Step Pressure Sensor Calibration Procedure
1 Assemble the pressure sensor calibration equipment as shown above, but DO
NOT connect the analyzer to the calibration equipment at this time.
2 If not already done, power ON the analyzer and display the CALIBRATION
menu. Note that this requires password validation (see page 55).
3 Use the UP () and DOWN () arrow buttons to select PRESSURE and then
press ENTER to display the CALIBRATE PRESSURE screen.
"Measured" is the pressure value currently being detected by the pressure
sensor, while "Applied” is a known value of pressure that will be applied for
calibration purposes.
4 With both the -ΔP and +ΔP ports open to the atmosphere, observe that the
current measured pressure reading should be 0.00 ± 0.01 inwc. If necessary,
zero the pressure sensor (Menu Pressure Zero) then repeat steps 2
through 4).
5 Connect the hose from the manometer to the +ΔP port and apply a negative
pressure to this port by adjusting the bellows for a manometer reading of
-4.00 (negative 4.00). Other units are shown below.
Units Name
Nominal Calibration
Point
inwc inches water column -4.00 inwc
mB millibars -10.00 mB
hPa hecto Pascals -10.00 hPa
Pa Pascals -1000 Pa
mm H
2
O millimeters of water -101.6 mm H
2
O
6 Use the UP (), DOWN (), LEFT (), and RIGHT () arrow buttons to enter
an “Applied” value that exactly equals the manometer reading.
NOTE: The calibration range is from -6 to -2 inwc (-15 to -5 mB).
An attempt to calibrate outside this range will cause the
message "Applied Value High" (or Low) to appear at the bottom
of the screen.
Przeglądanie stron 92
1 2 ... 88 89 90 91 92 93 94 95 96 97 98 ... 115 116

Komentarze do niniejszej Instrukcji

Brak uwag