7.3. Emptying the Water Trap Chamber
The water trap chamber should be emptied after every test, or when the water
condensate approaches the tip of the riser tube.
1. Remove water trap chamber (also see page 49).
2. Pour out liquid condensate and then reassemble trap.
7.4. Replacing the Filter Element
Replace the filter element when it becomes visibly dirty or becomes saturated
with water.
7.4.1. Material Required
Filter Element (P/N 0007-1644)
Small Flat Blade Screwdriver
7.4.2. Procedure
1. Remove water trap chamber (also see page 49).
2. Pry apart filter chamber using a small flat-blade screwdriver. Remove
and discard old filter.
3. Install new filter and reassemble filter chamber, making sure that
surfaces “A” and “B” contact each other.
4. Reassemble trap.
Figure 7-6. Disassembling the Water Trap/Filter Assembly