5.3. Low Battery Alarm
When the batteries are nearly depleted, an empty battery icon appears in the
upper right corner of the display, and a short beep is sounded every 10
seconds.
After a low battery alarm occurs, the analyzer will continue to operate for only a
few minutes. The amount of operating time that remains depends on many
factors (e.g., pump and keypad backlight being ON or OFF, and the type and
condition of the batteries).
5.4. Locating Sampling Points
Forced Air Furnace Hot Water Tank
Figure 5-2. Forced Air Furnace
Figure 5-3. Hot Water Tank
For atmospheric burner or gravity vented,
forced air heating equipment with a
clamshell or sectional heat exchanger
design, test each of the exhaust ports at
the top of the heat exchanger. The probe
should be inserted back into each of the
exhaust ports to obtain a flue gas sample,
before any dilution air is mixed in.
Domestic hot water tanks with the ‘bell’
shaped draft diverter can be accurately
tested by inserting the probe tip directly
into the top of the fire tube below the
diverter.