Bacharach also offers a convenient exchange program that allows the customer
to regularly send in old sensors and, as scheduled, receive calibrated
replacements. Contact Bacharach customer service for more details about the
exchange program.
6.4. Pressure Sensor Calibration
This procedure calibrates the pressure sensor to a known pressure value.
6.4.1. Material Required
•
Bellows
•
Manometer:
Range: ±8 in. of water column (±20 mb)
Accuracy: ±0.01 in. of water column (±0.025 mb)
6.4.2. Procedure
NOTE: The unit of measure for pressure is selected in the Pressure
Units Selection section on page 27
. In the following procedure, inwc
is selected, but note that any unit of measure can be used for
calibration purposes.
1. Assemble the pressure sensor calibration equipment as shown in
Figure 6-3, but DO NOT connect the analyzer to the calibration
equipment at this time.
2. If not already done, turn ON the analyzer and display the
CALIBRATION LIST screen (refer to Calibration on page 66).
3. Use the ▲▼ buttons to highlight Pressure, and then press ENT to
display the CALIBRATE PRESSURE screen.
NOTE: “Measured” is the pressure value currently being
detected by the pressure sensor, while “Applied” is a
known value of pressure that will be applied for calibration
purposes.
4. With both the –∆P and +∆P ports open to the atmosphere, observe that
the current Measured pressure reading should be 0 ±0.01  inwc. If
necessary, zero the pressure sensor (refer to Performing a Draft /
Pressure Measurement on page 51) and repeat Steps 2 thru 4.
5. Connect the hose from the manometer to the +∆P port and apply a
negative pressure to this port by adjusting the bellows for a manometer
reading of –4.00.
6. Use the ▲▼ and ◄► buttons to enter an applied value that exactly
equals the manometer reading.
Przeglądanie stron 67
1 2 ... 63 64 65 66 67 68 69 70 71 72 73 ... 103 104

Komentarze do niniejszej Instrukcji

Brak uwag

Ames-fire-waterworks 200B Bronze Double Check Valve Assemblies manuales

Manuales del propietario y guías del usuario para Equipo Ames-fire-waterworks 200B Bronze Double Check Valve Assemblies.
Ofrecemos 1 manuales en pdf Ames-fire-waterworks 200B Bronze Double Check Valve Assemblies para descargar gratis por tipos de documentos: Manual de usuario






Más productos y manuales para Equipo Ames-fire-waterworks

Modelos Tipo de documento
400B Bronze Reduced Pressure Zone Assemblies Manual de usuario   Ames Fire & Waterworks 400B Bronze Reduced Pressure Zone Assemblies User Manual, 8 paginas
3000B-FP Bronze Double Check Detector Assemblies w Manual de usuario   Ames Fire & Waterworks 3000B-FP Bronze Double Check Detector Assemblies with Gear Operated Slow Close Isolation Valves User Manual, 4 paginas
5000RPDA Reduced Pressure Detector Assemblies Manual de usuario   Ames Fire & Waterworks 5000RPDA Reduced Pressure Detector Assemblies User Manual, 16 paginas
4000CIV Cast Iron Reduced Pressure Zone Assemblies Manual de usuario   Ames Fire & Waterworks 4000CIV Cast Iron Reduced Pressure Zone Assemblies User Manual, 8 paginas
5001SS Reduced Pressure Detector Assemblies Manual de usuario   Ames Fire & Waterworks 5001SS Reduced Pressure Detector Assemblies User Manual, 4 paginas
5000CIV Cast Iron Reduced Pressure Detector Assemb Manual de usuario   Ames Fire & Waterworks 5000CIV Cast Iron Reduced Pressure Detector Assemblies User Manual, 8 paginas