PCA
3 Manual
32 0024-9473 Rev 4
4.11. Time Setup
The time is stored in the format: hh:mm:ss. Its value is part of the date and time
stamp that is saved along with each combustion test record. Set the analyzer’s
internal clock to the current time as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight TIME, and then press ENT to display
the TIME MENU.
4. First use the
◄►
buttons to move the cursor across the screen until it is
over the digit to be changed, and then press the ▲▼ buttons until the
desired value is displayed.
5. Repeat Step 4 until values for hour and minute are set.
NOTE: The value for seconds cannot be entered, but is
displayed and stored as part of the combustion test record.
6. Press ENT to save the displayed time values and re-display the SETUP
MENU, or press ESC to abort and retain the old time values.
Figure 4-11. Time Setup
4.12. O
2
Reference Setup
The measured values of CO, NOx, and SO
2
can be individually referenced to a
specific O
2
percentage of between 0 and 15%. Individually set up the O
2
reference value for each of the above gases as follows.
1. Display the MAIN MENU by pressing the MENU (F2) button. If
necessary, press ESC until MENU appears above F2.
2. Use the ▲▼ buttons to highlight SETUP, and then press ENT to display
the SETUP MENU.
3. Use the ▲▼ buttons to highlight O
2
REF, and then press ENT to
display the O
2
REFERENCE screen.