36
1-800-736-4666 • MyBacharach.com
Safety & Specialty Products
PGM-IR Portable Gas Analyzer
• Detection and measurement of CFC, HFC and HCFC
refrigerants, as well as CO
2
, N
2
O
and SF
6
gases
• Pinpoint low-level leaks with display and alarm
• Locates leaks in contaminated or high-ventilation areas
• Applications include refrigerant gas detection, trace gas
analysis and medical monitoring
• Two-year warranty
• Instrument comes with battery pack, probe, charger and
soft case with shoulder strap
• Up to 12 hours operation on fully charged battery
• LCD screen displays real-time gas concentrations
The PGM-IR is an advanced portable monitor that uses long-life non-
dispersive infrared (NDIR) sensor technology for analyzing and checking
low-level gas leaks. The unit can identify and pinpoint very small levels
of refrigerants, nitrous oxide or carbon dioxide gas in ambient air,
showing real-time concentrations in either parts per million (ppm) or
elapsed time weighted average (TWA) with a resolution down to 1 ppm.
The unit self-calibrates in 4 minute intervals by drawing ambient air
through its lter to provide a good “zero” baseline, enabling the monitor
to be used in contaminated environments without affecting the readings
of the target gas. Additional features include datalogging of up to
200 readings, and a leak check mode with "click-rate" audible / visual
alarms.
SENSING TECHNOLOGY
Non-Dispersive Infrared (NDIR)
GASES DETECTED
Refrigerants CFC, HFC and HCFC unit: R-11, R-12, R-21, R-22,
R-23, R-113, R-114, R-123, R-124, R-125, R-134A,
R-227, R-236FA, R-245FA, R-401A, R-402A, R-402B,
R-404A, R-407A, R-407C, R-408A, R-409A, R-410A,
R-422A, R-422D, R-424A, R-426A, R-427A, R-438A,
R-500, R-502, R-503, R-507, R-508B, FA188, FC72,
H-1211, H-1234YF, H-1301, H-2402, HFP, N-1230
Carbon Dioxide (CO
2
)
Nitrous Oxide (N
2
O)
Sulfur Hexaoride (SF
6
)
MEASUREMENT RANGE/ RESOLUTION
Refrigerant Gases 0-10,000 ppm displayed in 1 ppm increments
Carbon Dioxide (CO
2
) 0-100,000 ppm displayed in 1 ppm increments
Nitrous Oxide (N
2
O) 0-1,000 ppm displayed in 1 ppm increments
Sulfur Hexaoride (SF
6
) 0-1,000 ppm displayed in 1 ppm increments
RESPONSE TIME
90% of response within 5 seconds; 100% in
7 seconds
WARM UP TIME
CO
2
, Refrigerant gases: 5 minutes
N
2
O and SF
6
: 15 minutes
OTHER SPECIFICATIONS
Flow Rate 1 L/min typical
Power Li-Ion rechargeable battery with battery life indicator
Operating Temperature
Range 32 to 122°F (0 to 50°C)
Storage Temperature
Range –4 to 122°F (–20 to 50°C)
Operating Humidity
Range 5 to 90% RH, non condensing
Size 8”D x 19”W x 5”H (20.3 cm x 48.3 cm x 12.7 cm)
Weight 6 lbs. 12oz (3.07 kg)
Leak Check Mode Internal audible alarm, programmable
w/ Probe Visual alarm indicator
Warranty 2 Years (excluding contamination)
ORDERING INFORMATION
3015-5696 PGM-IR Analyzer: HFC, CFC, HCFC Refrigerant Gases
(without internal purge-air bladder)
3015-8001 PGM-IR Analyzer: Carbon Dioxide (CO
2
)
3015-4790 PGM-IR Analyzer: Nitrous Oxide (N
2
O)
3015-5148 PGM-IR Analyzer: Sulfur Hexaoride (SF
6
)*
3015-5739 Replacement Battery Pack
3015-5764 Replacement Battery Pack with Carry Case
*Special Order Only
Patent # 6,590,690
The PGM-IR is not designed or certied for intrinsic safety and may