CONSTRUCTION GA-40 T plus
12
G the gas sensors are working even when the analyser is switched off - so the battery should never be
disconnected or completely discharged.
G the analyser should not be switched off before all the flue gases have been removed from the
analyser gas circuit.
G the sensor chamber must be kept dry.
G the switched off analyser should be kept in a cool place, which increases the lifetime of the gas
sensors.
Gas Cell type Range [ppm]
O2 2FO 0 - 20.9 %
CO A3E/D 0 - 4000
CO 3F/D 0 - 20 000
CO 3M/F 0 - 10 %
NO 3NF/F 0 - 5 000
NO2 3ND 0 - 1000
SO2 3SF 0 - 5000
H2S 3H 0 - 1000
H2 3HYT 0 - 2000
HxCy PELLISTOR 0 - 3 %
CO2 IR 0 - 50 %
Tab. 1 Available measuring cells and ranges
4.2.2 Ambient air temperature sensor
The ambient temperature is measured by the analyser for use in some of the calculations carried out, such as
Stack Loss. This can be carried out either by the temperature compensation sensor built into the plug of the
probe holder (INT) or the external connector T3 (EXT). The temperature sensor is chosen under the suboption
OPTIONS / SENSORS / ANALOGUE INPUTS.
4.2.3 Flue gas temperature sensor
Flue gas temperature sensor is situated in the gas probe pipe and plugged into PROBE socket on the front
panel of the analyser. This sensor is a NiCr-Ni thermocouple in the standard probe version. The range of such
a thermocouple is 50-800
o
C in the standard version (1100°C available optionally). A special gas probe with
PtRh-Pt thermocouple can also be used, with a range up to 1600
o
C. Proper flue gas temperature measurement
is the most important factor for correct calculation of the chimney loss and other heat engineering results.
4.2.4 Gas chamber temperature sensor
To compensate for temperature drift of the gas cells, the temperature of the gas chamber is constantly
monitored.
4.2.5 Differential pressure sensor
The GA-40 T plusplus is fitted with a precise semiconductor differential pressure sensor. Stubs marked +/-
PRESSURE situated on the left side of the front panel are used to measure pressure. Negative pressure in flue
gas channel (chimney draught) can be measured with the one stub only; differential pressure measurement is
possible using both stubs. The measuring range of the pressure sensor is +/-50 hPa. This sensor can also be
used for measurement of the flow rate in the stack.
4.2.6 Analogue inputs
The six analogue inputs installed can be used to measure voltage or current on two channels (U/I1 and U/I2),
temperature using thermocouples (T1 EXT and T2 EXT) and temperature using thermistors (T3 EXT and T4
EXT). The values measured can be viewed on the screen or stored in the EEPROM memory.